Moisture Analyzers
A moisture analyzer is a precision balance combined with a halogen or infrared heater that reports loss on drying as a percentage of the initial sample mass. It replaces an oven-and-desiccator procedure that takes hours with a measurement that takes minutes.
Also called: moisture balance · loss on drying analyzer · halogen moisture analyzer · LOD analyzer
Moisture Analyzers in stock
Shipped from the United States with the manufacturer's warranty. Prices shown are current US pricing; call 832-290-3120 for volume quotes or application help.
How moisture analyzers work and how to specify them
Results are method-dependent, which surprises new users. Sample mass, particle size, heating profile and end-point criterion all change the number, so a moisture analyzer result is only comparable to another result produced with the same method. Standardize a method per material and store it as a program.
Sample preparation matters more than the instrument. Spread the sample evenly and thinly across the pan — a mound dries from the outside in and terminates early. Use glass-fiber pads for liquids and pastes to increase surface area.
For regulated QC work, correlate your rapid method against the reference oven method once, document the offset, and re-verify whenever the material specification changes. Every instrument can ship with a NIST-traceable calibration certificate performed in our Cypress, Texas service lab, and we handle on-site calibration and repair across the Houston metro.
Moisture analyzer selection
| Application | Typical capacity | Typical readability | Notes |
|---|---|---|---|
| Food QC, snacks, dairy | 50–110 g | 0.001 g / 0.01% | Halogen heating |
| Plastics and resin pellets | 50–110 g | 0.001 g | High-temperature program |
| Grain and agriculture | 110–200 g | 0.01 g | Larger pan for bulk samples |
| Pharma and fine chemicals | 50 g | 0.0001 g / 0.001% | High-resolution, GLP output |
What to check before you buy
- Readability in % moisture. 0.01% is standard; 0.001% only matters for low-moisture pharma samples.
- Heating element. Halogen heats fastest and most evenly; infrared and metal-rod elements are more tolerant of splatter.
- Method memory. Stored programs keep operators from changing the result by changing the procedure.
- End-point criteria. Automatic, timed and rate-based end points each suit different materials.
- Data output. GLP printout or PC output for QC records.
Moisture Analyzers buyer's guide: how to choose the right model
Work through these criteria in order. Each one narrows the shortlist, and getting them in this sequence is what prevents an expensive mismatch between the instrument and the job.
1. Confirm moisture analyzers are the right class
You need loss-on-drying moisture content, not just weight. Balance with halogen or infrared heating chamber. If that does not describe your job, the wrong class will cost you twice — plain weighing tasks — the heater adds cost you will not use.
Key criterion 1
Match sample size and pan diameter to your material; thin even layers dry repeatably.
Key criterion 2
Halogen heats faster than infrared; both need a validated drying profile per material.
Key criterion 3
Store drying programs in memory so operators cannot vary the method.
Key criterion 4
Budget for spare sample pans and glass fibre filters as consumables.
2. Size the capacity with headroom
This class runs 50 g – 200 g sample. Work from your heaviest realistic load including packaging, pallet or container, then step up to the next capacity rather than running an instrument at its ceiling every day.
3. Set readability from the smallest thing you measure
Typical readability here is 0.001 g / 0.01% moisture. Choose resolution from the smallest increment you must resolve — not the largest weight — and remember that finer resolution means a slower, more environment-sensitive reading.
4. Match the instrument to the environment
Food, plastics, pharma and agricultural QC labs. Decide on washdown rating, mounting and display position before comparing prices; an under-specified enclosure is the most common reason a scale is replaced early.
5. Decide whether it must be legal for trade
Not applicable. If money changes hands on the basis of the weight, you need an approved model sealed by your state weights-and-measures authority; if it is an internal process check, you do not.
6. Plan data, service and calibration
Choose the data path you will actually use — printout, serial, USB or PLC — and budget annual calibration. We calibrate, certify and repair moisture analyzers in our Cypress, Texas lab and on site across the Houston metro.
Moisture Analyzers vs Lab Balances vs Analytical Balances vs Industrial Scales
These categories overlap in capacity but not in how the load is presented. Compare the rows below, then follow the link to the category that matches your application.
| Category | How it takes the load | Typical capacity | Typical readability | Use cases | Choose it when | Legal for trade |
|---|---|---|---|---|---|---|
| Moisture Analyzers (this page) | Balance with halogen or infrared heating chamber | 50 g – 200 g sample | 0.001 g / 0.01% moisture | Food, plastics, pharma and agricultural QC labs | You need loss-on-drying moisture content, not just weight | Not applicable |
| Lab Balances | Bench balance with draft shield on finer models | 120 g – 8 kg | 0.001 g – 0.1 g | QC labs, pharmacies, universities, production labs | You need milligram or centigram resolution with GLP output | NTEP versions available where sales by weight occur |
| Analytical Balances | Enclosed balance with glass draft chamber | 80 g – 320 g | 0.0001 g (0.1 mg) | Analytical and research labs, marble or granite benches | Method or pharmacopoeia requires 0.1 mg resolution | Not applicable — laboratory measurement, not trade |
| Industrial Scales | Any rugged production-floor weighing platform or system | 50 lb – 20,000 lb | 0.1 lb – 5 lb | Manufacturing, recycling, agriculture, logistics | The scale lives on a production floor and must survive it | NTEP available across most industrial classes |
Certifications, datasheets and documentation
What we can certify for moisture analyzers, and the documents you can download before you raise a purchase order.
Authorized A&D Weighing dealer
Sold as an authorized dealer since 1998, so every unit carries the full manufacturer warranty and factory firmware — not grey-market stock.
NIST-traceable calibration certificate
Any instrument can ship with a calibration certificate produced in our Cypress, Texas lab using mass standards traceable to NIST, with as-found and as-left values recorded.
In-house repair & on-site service
Factory-trained technicians repair, adjust and re-certify weighing equipment in our own Texas workshop and on-site across the Houston metro.
Moisture Analyzers — specification datasheet (PDF)
Capacity and readability selection table, buying criteria, certification notes and the answers to the questions purchasing always asks — two pages, no email required.
Download datasheetInstruction manuals hosted here
- A&D FX-i / FZ-i Series instruction manualCovers: FX-120i, FX-123, FX-200i, FX-300i, FX-500i, FX-1200i, FX-2000i, FX-3000i, FZ-120i–FZ-3000i
- A&D GX Series instruction manualCovers: GX-200, GX-400, GX-600, GX-800, GX-1000, GX-2000, GX-4000, GX-6100, GX-6000, GX-8000, GX-10K
- A&D GF Series instruction manualCovers: GF-200, GF-300, GF-400, GF-600, GF-1000, GF-2000, GF-3000, GF-4000, GF-6100, GF-8000, GF-10K
Moisture Analyzers glossary — the specification terms explained
Every term used in the tables above, defined the way NIST Handbook 44, ASTM E617 and OIML R76/R111 use it.
- Accuracy
- How close a displayed weight is to the sample's true mass. Accuracy is limited by readability, repeatability and linearity together, so a scale can never be more accurate than its specification sheet's worst of those three figures.
- Analytical balance
- A balance reading to 0.0001 g (0.1 mg) or finer, enclosed in a draft chamber. Used for analytical chemistry, pharmaceutical assay and trace-level formulation.
- ASTM E617
- The American standard that defines tolerance classes for mass standards. Class 1 weights are the tightest commonly used for laboratory balances, Class 2 for 0.01 g balances, and Classes 4 and 6 for industrial platform and crane scales.
- Auto-zero tracking
- A function that continuously re-zeros small drifts near zero so the display stays at 0 when the pan is empty. It can mask a very slow addition of material, so it is disabled for some dispensing work.
- Calibration
- Adjusting a scale's span so it reads a certified test weight correctly, documented with as-found and as-left values. Distinct from verification, which records a reading without changing anything.
- Drift
- A slow change in reading with no change in load, usually caused by temperature change, air movement, static, or a settling mechanical assembly rather than by an instrument fault.
- Force restoration
- A weighing mechanism that balances the load with an electromagnetic coil and measures the current needed to do so. It gives faster settling, better linearity and lower drift than a strain gauge, and is what A&D's SHS sensor uses.
- GLP output
- A printout or data record that includes date, time, instrument identification and calibration status, as required by Good Laboratory Practice and comparable quality systems for audit trails.
- Moisture analyzer
- A balance combined with a controlled heater that weighs a sample, dries it, and reports moisture content as a percentage of the mass lost. Also called a loss-on-drying balance.
- Precision balance
- A balance reading to between 0.1 g and 0.001 g, usually with an open pan. Coarser than an analytical balance and far finer than a bench or platform scale.
- Resolution, internal
- The number of counts the analog-to-digital converter actually resolves internally, typically ten to a hundred times finer than the displayed readability. High internal resolution is what makes stable, fast display averaging possible.
- Stabilisation time
- How long after a load is applied before the reading is stable enough to record. It determines real throughput far more than capacity does in high-volume weighing.
Moisture Analyzers — frequently asked questions
- How does a moisture analyzer work?
- It weighs the sample, heats it with a halogen or infrared element while continuously re-weighing, and reports the percentage of mass lost as moisture once the weight stabilizes.
- Is a moisture analyzer as accurate as an oven?
- It is accurate and repeatable, but it measures loss on drying rather than water specifically. Correlate the rapid method to your oven reference method once per material and document the offset.
- What sample size should I use?
- Most methods use 2–10 g spread thinly and evenly across the pan. Larger samples take longer; uneven samples terminate early and read low.
- What temperature should I set?
- It depends on the material — typically 105 °C for food and grain, higher for plastics. Start from the reference oven method's temperature and validate against it.