Moisture Analyzers

    A moisture analyzer is a precision balance combined with a halogen or infrared heater that reports loss on drying as a percentage of the initial sample mass. It replaces an oven-and-desiccator procedure that takes hours with a measurement that takes minutes.

    Also called: moisture balance · loss on drying analyzer · halogen moisture analyzer · LOD analyzer

    Moisture Analyzers in stock

    Shipped from the United States with the manufacturer's warranty. Prices shown are current US pricing; call 832-290-3120 for volume quotes or application help.

    Loading moisture analyzers…

    How moisture analyzers work and how to specify them

    Results are method-dependent, which surprises new users. Sample mass, particle size, heating profile and end-point criterion all change the number, so a moisture analyzer result is only comparable to another result produced with the same method. Standardize a method per material and store it as a program.

    Sample preparation matters more than the instrument. Spread the sample evenly and thinly across the pan — a mound dries from the outside in and terminates early. Use glass-fiber pads for liquids and pastes to increase surface area.

    For regulated QC work, correlate your rapid method against the reference oven method once, document the offset, and re-verify whenever the material specification changes. Every instrument can ship with a NIST-traceable calibration certificate performed in our Cypress, Texas service lab, and we handle on-site calibration and repair across the Houston metro.

    Moisture analyzer selection

    Moisture analyzer selection
    ApplicationTypical capacityTypical readabilityNotes
    Food QC, snacks, dairy50–110 g0.001 g / 0.01%Halogen heating
    Plastics and resin pellets50–110 g0.001 gHigh-temperature program
    Grain and agriculture110–200 g0.01 gLarger pan for bulk samples
    Pharma and fine chemicals50 g0.0001 g / 0.001%High-resolution, GLP output

    What to check before you buy

    • Readability in % moisture. 0.01% is standard; 0.001% only matters for low-moisture pharma samples.
    • Heating element. Halogen heats fastest and most evenly; infrared and metal-rod elements are more tolerant of splatter.
    • Method memory. Stored programs keep operators from changing the result by changing the procedure.
    • End-point criteria. Automatic, timed and rate-based end points each suit different materials.
    • Data output. GLP printout or PC output for QC records.

    Moisture Analyzers buyer's guide: how to choose the right model

    Work through these criteria in order. Each one narrows the shortlist, and getting them in this sequence is what prevents an expensive mismatch between the instrument and the job.

    1. 1. Confirm moisture analyzers are the right class

      You need loss-on-drying moisture content, not just weight. Balance with halogen or infrared heating chamber. If that does not describe your job, the wrong class will cost you twice — plain weighing tasks — the heater adds cost you will not use.

    2. Key criterion 1

      Match sample size and pan diameter to your material; thin even layers dry repeatably.

    3. Key criterion 2

      Halogen heats faster than infrared; both need a validated drying profile per material.

    4. Key criterion 3

      Store drying programs in memory so operators cannot vary the method.

    5. Key criterion 4

      Budget for spare sample pans and glass fibre filters as consumables.

    6. 2. Size the capacity with headroom

      This class runs 50 g – 200 g sample. Work from your heaviest realistic load including packaging, pallet or container, then step up to the next capacity rather than running an instrument at its ceiling every day.

    7. 3. Set readability from the smallest thing you measure

      Typical readability here is 0.001 g / 0.01% moisture. Choose resolution from the smallest increment you must resolve — not the largest weight — and remember that finer resolution means a slower, more environment-sensitive reading.

    8. 4. Match the instrument to the environment

      Food, plastics, pharma and agricultural QC labs. Decide on washdown rating, mounting and display position before comparing prices; an under-specified enclosure is the most common reason a scale is replaced early.

    9. 5. Decide whether it must be legal for trade

      Not applicable. If money changes hands on the basis of the weight, you need an approved model sealed by your state weights-and-measures authority; if it is an internal process check, you do not.

    10. 6. Plan data, service and calibration

      Choose the data path you will actually use — printout, serial, USB or PLC — and budget annual calibration. We calibrate, certify and repair moisture analyzers in our Cypress, Texas lab and on site across the Houston metro.

    Moisture Analyzers vs Lab Balances vs Analytical Balances vs Industrial Scales

    These categories overlap in capacity but not in how the load is presented. Compare the rows below, then follow the link to the category that matches your application.

    Moisture Analyzers vs Lab Balances vs Analytical Balances vs Industrial Scales — selection comparison
    CategoryHow it takes the loadTypical capacityTypical readabilityUse casesChoose it whenLegal for trade
    Moisture Analyzers (this page)Balance with halogen or infrared heating chamber50 g – 200 g sample0.001 g / 0.01% moistureFood, plastics, pharma and agricultural QC labsYou need loss-on-drying moisture content, not just weightNot applicable
    Lab BalancesBench balance with draft shield on finer models120 g – 8 kg0.001 g – 0.1 gQC labs, pharmacies, universities, production labsYou need milligram or centigram resolution with GLP outputNTEP versions available where sales by weight occur
    Analytical BalancesEnclosed balance with glass draft chamber80 g – 320 g0.0001 g (0.1 mg)Analytical and research labs, marble or granite benchesMethod or pharmacopoeia requires 0.1 mg resolutionNot applicable — laboratory measurement, not trade
    Industrial ScalesAny rugged production-floor weighing platform or system50 lb – 20,000 lb0.1 lb – 5 lbManufacturing, recycling, agriculture, logisticsThe scale lives on a production floor and must survive itNTEP available across most industrial classes

    Certifications, datasheets and documentation

    What we can certify for moisture analyzers, and the documents you can download before you raise a purchase order.

    • Authorized A&D Weighing dealer

      Sold as an authorized dealer since 1998, so every unit carries the full manufacturer warranty and factory firmware — not grey-market stock.

    • NIST-traceable calibration certificate

      Any instrument can ship with a calibration certificate produced in our Cypress, Texas lab using mass standards traceable to NIST, with as-found and as-left values recorded.

    • In-house repair & on-site service

      Factory-trained technicians repair, adjust and re-certify weighing equipment in our own Texas workshop and on-site across the Houston metro.

    Moisture Analyzers — specification datasheet (PDF)

    Capacity and readability selection table, buying criteria, certification notes and the answers to the questions purchasing always asks — two pages, no email required.

    Download datasheet

    Instruction manuals hosted here

    All scale manuals and literature

    Moisture Analyzers glossary — the specification terms explained

    Every term used in the tables above, defined the way NIST Handbook 44, ASTM E617 and OIML R76/R111 use it.

    Accuracy
    How close a displayed weight is to the sample's true mass. Accuracy is limited by readability, repeatability and linearity together, so a scale can never be more accurate than its specification sheet's worst of those three figures.
    Analytical balance
    A balance reading to 0.0001 g (0.1 mg) or finer, enclosed in a draft chamber. Used for analytical chemistry, pharmaceutical assay and trace-level formulation.
    ASTM E617
    The American standard that defines tolerance classes for mass standards. Class 1 weights are the tightest commonly used for laboratory balances, Class 2 for 0.01 g balances, and Classes 4 and 6 for industrial platform and crane scales.
    Auto-zero tracking
    A function that continuously re-zeros small drifts near zero so the display stays at 0 when the pan is empty. It can mask a very slow addition of material, so it is disabled for some dispensing work.
    Calibration
    Adjusting a scale's span so it reads a certified test weight correctly, documented with as-found and as-left values. Distinct from verification, which records a reading without changing anything.
    Drift
    A slow change in reading with no change in load, usually caused by temperature change, air movement, static, or a settling mechanical assembly rather than by an instrument fault.
    Force restoration
    A weighing mechanism that balances the load with an electromagnetic coil and measures the current needed to do so. It gives faster settling, better linearity and lower drift than a strain gauge, and is what A&D's SHS sensor uses.
    GLP output
    A printout or data record that includes date, time, instrument identification and calibration status, as required by Good Laboratory Practice and comparable quality systems for audit trails.
    Moisture analyzer
    A balance combined with a controlled heater that weighs a sample, dries it, and reports moisture content as a percentage of the mass lost. Also called a loss-on-drying balance.
    Precision balance
    A balance reading to between 0.1 g and 0.001 g, usually with an open pan. Coarser than an analytical balance and far finer than a bench or platform scale.
    Resolution, internal
    The number of counts the analog-to-digital converter actually resolves internally, typically ten to a hundred times finer than the displayed readability. High internal resolution is what makes stable, fast display averaging possible.
    Stabilisation time
    How long after a load is applied before the reading is stable enough to record. It determines real throughput far more than capacity does in high-volume weighing.
    Full scale glossary — every weighing term defined

    Moisture Analyzers — frequently asked questions

    How does a moisture analyzer work?
    It weighs the sample, heats it with a halogen or infrared element while continuously re-weighing, and reports the percentage of mass lost as moisture once the weight stabilizes.
    Is a moisture analyzer as accurate as an oven?
    It is accurate and repeatable, but it measures loss on drying rather than water specifically. Correlate the rapid method to your oven reference method once per material and document the offset.
    What sample size should I use?
    Most methods use 2–10 g spread thinly and evenly across the pan. Larger samples take longer; uneven samples terminate early and read low.
    What temperature should I set?
    It depends on the material — typically 105 °C for food and grain, higher for plastics. Start from the reference oven method's temperature and validate against it.