Calibration Test Weights

    A calibration test weight is a certified mass used to verify or adjust a scale. Weights are graded by ASTM E617 (Class 0 to 7) or OIML R111 (E1 to M3), and the working rule is that the weight's tolerance should be no more than one third of the balance's permitted error at that load.

    Also called: test weight · ASTM weight · OIML weight · check weight · calibration mass

    Calibration Weights in stock

    Shipped from the United States with the manufacturer's warranty. Prices shown are current US pricing; call 832-290-3120 for volume quotes or application help.

    Loading calibration weights…

    How calibration weights work and how to specify them

    Choosing a class is a straightforward calculation. Take the balance's readability, multiply by the number of divisions of permitted error at the test load, divide by three, and select the class whose tolerance at that nominal value is smaller. In practice: ASTM Class 1 for analytical balances, Class 2 for precision balances, Class 3–4 for bench scales and Class 5–6 for floor and industrial scales.

    Handling determines whether the weight keeps its class. Never touch a Class 1–3 weight with bare hands — skin oils add measurable mass. Use the supplied forceps or gloves, store in the fitted case, and allow the weight to equilibrate to room temperature before use.

    Certified weights require periodic recertification, typically annually for working standards, because wear and handling shift mass over time. We supply weights with NIST-traceable certificates and recertify customer weights in our Cypress, Texas lab.

    Which weight class for which instrument

    Which weight class for which instrument
    ApplicationTypical capacityTypical readabilityNotes
    Analytical balances (0.1 mg)1 mg – 200 gASTM Class 1OIML E2 equivalent
    Precision balances (0.001–0.01 g)1 g – 2 kgASTM Class 2OIML F1 equivalent
    Bench and counting scales100 g – 20 kgASTM Class 3–4OIML F2/M1 equivalent
    Floor, crane and industrial scales5 – 50 lb+ASTM Class 5–6Cast iron / stainless field weights

    What to check before you buy

    • One-third rule. Weight tolerance ≤ 1/3 of the balance's permitted error at that load.
    • Material. Stainless for Class 1–4; cast iron is acceptable for field-class industrial verification.
    • Nominal value. Verify near full capacity, plus a mid-scale point for linearity.
    • Certificate. NIST-traceable, with as-found and as-left values for regulated use.
    • Recertification interval. Annually for working standards; sooner if handled frequently.

    Calibration Weights buyer's guide: how to choose the right model

    Work through these criteria in order. Each one narrows the shortlist, and getting them in this sequence is what prevents an expensive mismatch between the instrument and the job.

    1. 1. Confirm calibration weights are the right class

      You must calibrate or verify a balance or scale. Individual masses or sets, OIML/ASTM classified. If that does not describe your job, the wrong class will cost you twice — nothing — pair a weight class to each instrument you own.

    2. Key criterion 1

      Choose weight class from the instrument: roughly 1/3 of the scale's tolerance at the test point.

    3. Key criterion 2

      Test at span near capacity, not just at a low point, and re-zero between checks.

    4. Key criterion 3

      Order a traceable calibration certificate with the weights, then re-certify on a schedule.

    5. Key criterion 4

      Handle with gloves or forceps; fingerprints and corrosion change mass.

    6. 2. Size the capacity with headroom

      This class runs 1 mg – 25 kg per piece. Work from your heaviest realistic load including packaging, pallet or container, then step up to the next capacity rather than running an instrument at its ceiling every day.

    7. 3. Set readability from the smallest thing you measure

      Typical readability here is Tolerance defined by class (E2, F1, M1 / ASTM 1–4). Choose resolution from the smallest increment you must resolve — not the largest weight — and remember that finer resolution means a slower, more environment-sensitive reading.

    8. 4. Match the instrument to the environment

      Every site that owns a scale. Decide on washdown rating, mounting and display position before comparing prices; an under-specified enclosure is the most common reason a scale is replaced early.

    9. 5. Decide whether it must be legal for trade

      Class selection follows the instrument's division count. If money changes hands on the basis of the weight, you need an approved model sealed by your state weights-and-measures authority; if it is an internal process check, you do not.

    10. 6. Plan data, service and calibration

      Choose the data path you will actually use — printout, serial, USB or PLC — and budget annual calibration. We calibrate, certify and repair calibration weights in our Cypress, Texas lab and on site across the Houston metro.

    Calibration Weights vs Analytical Balances vs Lab Balances vs NTEP Scales

    These categories overlap in capacity but not in how the load is presented. Compare the rows below, then follow the link to the category that matches your application.

    Calibration Weights vs Analytical Balances vs Lab Balances vs NTEP Scales — selection comparison
    CategoryHow it takes the loadTypical capacityTypical readabilityUse casesChoose it whenLegal for trade
    Calibration Weights (this page)Individual masses or sets, OIML/ASTM classified1 mg – 25 kg per pieceTolerance defined by class (E2, F1, M1 / ASTM 1–4)Every site that owns a scaleYou must calibrate or verify a balance or scaleClass selection follows the instrument's division count
    Analytical BalancesEnclosed balance with glass draft chamber80 g – 320 g0.0001 g (0.1 mg)Analytical and research labs, marble or granite benchesMethod or pharmacopoeia requires 0.1 mg resolutionNot applicable — laboratory measurement, not trade
    Lab BalancesBench balance with draft shield on finer models120 g – 8 kg0.001 g – 0.1 gQC labs, pharmacies, universities, production labsYou need milligram or centigram resolution with GLP outputNTEP versions available where sales by weight occur
    NTEP ScalesAny class of scale holding a Certificate of ConformanceDepends on classClass II/III division limits applyAnywhere goods change hands by weightYou buy, sell or bill by weightThis is the category — always verify the CoC number

    Certifications, datasheets and documentation

    What we can certify for calibration weights, and the documents you can download before you raise a purchase order.

    • Authorized A&D Weighing dealer

      Sold as an authorized dealer since 1998, so every unit carries the full manufacturer warranty and factory firmware — not grey-market stock.

    • NIST-traceable calibration certificate

      Any instrument can ship with a calibration certificate produced in our Cypress, Texas lab using mass standards traceable to NIST, with as-found and as-left values recorded.

    • ASTM E617 / OIML R111 test weights

      Verification weights are supplied in ASTM Class 1–6 or OIML E2–M3 tolerances, sized so the standard's tolerance stays inside one third of the scale's smallest division.

    • In-house repair & on-site service

      Factory-trained technicians repair, adjust and re-certify weighing equipment in our own Texas workshop and on-site across the Houston metro.

    Calibration Weights — specification datasheet (PDF)

    Capacity and readability selection table, buying criteria, certification notes and the answers to the questions purchasing always asks — two pages, no email required.

    Download datasheet

    Instruction manuals hosted here

    Tell us the model and we will send the manufacturer's manual and declaration of conformity — call 832-290-3120.

    All scale manuals and literature

    Calibration Weights glossary — the specification terms explained

    Every term used in the tables above, defined the way NIST Handbook 44, ASTM E617 and OIML R76/R111 use it.

    Accuracy
    How close a displayed weight is to the sample's true mass. Accuracy is limited by readability, repeatability and linearity together, so a scale can never be more accurate than its specification sheet's worst of those three figures.
    ASTM E617
    The American standard that defines tolerance classes for mass standards. Class 1 weights are the tightest commonly used for laboratory balances, Class 2 for 0.01 g balances, and Classes 4 and 6 for industrial platform and crane scales.
    Calibration
    Adjusting a scale's span so it reads a certified test weight correctly, documented with as-found and as-left values. Distinct from verification, which records a reading without changing anything.
    Capacity
    The maximum mass a scale is rated to weigh. Usable capacity is the rated capacity minus every container, fixture or platform that sits on the pan, because tare does not add capacity back.
    Division (d)
    The smallest increment the display can change by, also called readability or scale interval. A 1200 g x 0.01 g balance has a 0.01 g division and 120,000 divisions across its range.
    Linearity
    The maximum deviation from a straight line between zero and full capacity, expressed in divisions. A balance can be perfectly repeatable and still be non-linear, which is why calibration is checked at more than one point.
    NIST Handbook 105-1
    The United States specification for field standard mass standards, and the source of the widely used rule that a test weight's tolerance should not exceed one third of the scale's smallest division.
    NIST traceable
    A calibration performed with standards whose values link through an unbroken chain of comparisons, each with stated uncertainty, back to national standards held by NIST.
    OIML R76 / R111
    The international recommendations for non-automatic weighing instruments (R76) and for weights of classes E1 to M3 (R111). They are the international counterparts to NIST Handbook 44 and ASTM E617.
    Readability
    The smallest increment a display can show. Readability is not accuracy: it sets the finest digit you can read, while accuracy also depends on repeatability and linearity.
    Repeatability
    The standard deviation of repeated weighings of the same load under identical conditions, quoted in grams or divisions. It sets the practical floor on how small a difference you can trust.
    Span
    The difference in reading between zero and full capacity. Span error is proportional to load, which is why verification must be done at or near full capacity to detect it.
    Full scale glossary — every weighing term defined

    Calibration Weights — frequently asked questions

    What class calibration weight do I need?
    Choose a class whose tolerance is no more than one third of your balance's permitted error at the test load — typically ASTM Class 1 for analytical balances, Class 2 for precision balances, and Class 3–6 for bench and industrial scales.
    What is the difference between ASTM and OIML weight classes?
    They are parallel systems. ASTM E617 runs Class 0 through 7; OIML R111 runs E1 through M3. Rough equivalents: ASTM 1 ≈ OIML E2, ASTM 2 ≈ F1, ASTM 4 ≈ F2, ASTM 5 ≈ M1.
    How often do test weights need recertification?
    Annually for weights in regular use, and immediately after any drop, corrosion or bare-hand contact on a Class 1–3 weight.
    Can I calibrate a scale without certified weights?
    You can adjust it, but the result is not traceable and will not satisfy an auditor, a state inspector or a quality system. Verification requires a certified mass.