Back to all A&D scales
    Radwag Moisture Balance PM 200/3Y Touch Screen 200g x 1mg,Touch screen,Brand New

    Radwag Moisture Balance PM 200/3Y Touch Screen 200g x 1mg,Touch screen,Brand New

    $2440.00

    $1899.00

    On Sale

    MSRP $2440.00 — you save $541.00 (22% off)

    Capacity 200 g × readability 1 mg.

    Radwag PM 200.3Y Moisture Analyzer – 200g x 0.001 gram, 2 year warranty, made in Europe

    Moisture analyzer is a laboratory measuring instrument intended to determine relative moisture content in samples of different substances. PM 3Y features 5,7″ LCD colour touch panel which provides new possibility within the scope of balance operation and presentation of measurement result. It features extended databases containing programmable drying modes related to the database of samples. The PM 3Y series enables printing and exporting charts presented on its display to a BMP file.

    Customization of instrument’s settings is possible thanks to extended user profiles. The balance level is monitored by LevelSENSING system, RADWAG patented solution, which uses a system of an electronic level. Standard and user defined printouts allow for maintaining documentation complying with GLP/GMP requirements practically in any organization.

    Max capacity [Max]

    200 g

    Readability [d]

    1 mg

    Tare range

    -200 g

    Pan Size

    ø 90, h= 8 mm

    Working temperature

    +10 – +40 °C

    Power supply

    100 ÷ 240 V AC 50 ÷ 60 Hz

    Display

    5,7” (touch screen)

    Interface

    2 x USB 2.0; 4IN / 4OUT; RS 232; Ethernet 10/100Mbit

    Packaging size

    595×395×420 mm

    Net weight/Gross weight

    6/10 kg

    Max sample weight

    200 g

    Moisture readout accuracy

    0,001 %

    Moisture content repeatability

    +/-0,05% (sample ~ 2g), +/-0,01% (sample ~ 10g)

    Drying temperature range

    max 160 °C

    Max sample height

    20 mm

    Heating module

    IR emiter

    Power of heating device

    400 W

    Drying mode

    4 drying modes (standard, quick, step, mild)

    Auto switch-off options

    4 modes (time, defined, automatic, manual)

    Additional functions

    sample identification, drying diagram

    Why buy from EK1200i.com

    • Authorized dealer since 1998 — new, factory-sealed equipment with the full manufacturer warranty
    • Free fast shipping within the mainland USA
    • In-house NIST-traceable calibration available before shipment on request
    • Live phone and email support from working scale technicians in Cypress, Texas
    • Full-service calibration and repair lab for the life of your scale

    Specifications are published by the manufacturer and may change without notice. Contact us if you need a current datasheet or help matching a model to your application.

    Request Quote
    Authorized A&D Dealer
    Free Mainland US Shipping
    Expert Support

    Need bulk pricing? Contact us for special pricing on volume orders.

    Specifications

    Manufacturer
    Radwag
    Model
    PM-200
    Instrument class
    Moisture analyzer
    Capacity
    200 g
    Readability (d)
    1 g
    Display divisions
    200
    SKU / Part number
    RADWA-PM200-3Y-200G-1-2
    Price (USD)
    $1899.00
    Availability
    In stock — ships from the USA
    Shipping
    Free ground shipping, mainland USA
    Warranty
    Full manufacturer warranty, registered to the buyer
    Calibration
    NIST-traceable, performed in Cypress, TX

    Capacity in other units

    Grams (g)
    200
    Kilograms (kg)
    0.2
    Pounds (lb)
    0.441
    Ounces (oz)
    7.05
    Troy ounces (ozt)
    6.43
    Carats (ct)
    1,000

    What is a moisture analyzer, and how does this one work?

    A moisture analyzer determines moisture content by loss on drying: it weighs a sample, heats it with a halogen or infrared element, and reports the percentage of mass lost as the sample dries to constant weight. The PM-200 handles samples up to 200 g with 1 g weighing readability. The result is a moisture percentage in minutes rather than the hours an oven method requires.

    The instrument is a precision balance with a heating chamber above the pan. It records initial wet mass, ramps the heater to a set temperature, and samples mass continuously. A switch-off criterion — fixed time, or a rate threshold such as less than 1 mg change per 50 seconds — ends the test. Moisture % = (wet mass − dry mass) ÷ wet mass × 100.

    Where the PM-200 is used

    • •Food processing: grain, flour, dairy, snack, and meat moisture QC
    • •Plastics: resin and pellet moisture before injection moulding
    • •Pharmaceutical granulation and drying endpoint control
    • •Agriculture: hay, feed, and seed moisture
    • •Paper, textile, and wood products

    Getting accurate results

    • •Spread the sample thin and evenly across the pan — 5 g in a uniform layer dries far more reproducibly than a 10 g mound.
    • •Use disposable glass-fibre pads for liquids and pastes to increase surface area.
    • •Run each new material three times to establish repeatability before adopting a method.
    • •Record temperature, sample mass, and switch-off criterion with every result; a moisture figure without its method is not comparable to anything.

    Frequently asked questions about the PM-200

    What are the capacity and readability of the PM-200?

    The PM-200 is rated to 200 g with 1 g readability, which is 200 display divisions across the range. Subtract the mass of any container or fixture from capacity to get your usable weighing range, because tare does not add capacity back.

    What test weight is needed to calibrate the PM-200?

    A certified weight at or near full capacity with a tolerance no greater than one third of the smallest division (1 g here), per NIST Handbook 105-1. That is ASTM E617 Class 1 or Class 2 for laboratory resolutions and Class 4 or 6 for industrial capacities. We stock certified weights and can supply the matching class with the instrument.

    How often should the PM-200 be calibrated?

    Annually for general commercial and industrial use, semi-annually or quarterly under ISO 9001, GMP or HACCP, and immediately after any relocation, suspected overload or impact. Legal-for-trade devices additionally follow the inspection cycle set by the local weights-and-measures jurisdiction.

    Is the PM-200 legal for trade?

    A scale that determines the price a customer pays must carry an NTEP Certificate of Conformance and be sealed by your state or local weights-and-measures authority. Laboratory-grade accuracy alone does not confer legal-for-trade status, and certification is lost if the seal is broken or the device is moved without re-verification. Ask us to confirm the NTEP number before ordering if money will change hands based on the weight.

    What is the difference between readability and accuracy?

    Readability is the smallest increment the display can show — 1 g on the PM-200. Accuracy is how close a reading is to true mass and is limited by readability, repeatability (the standard deviation of repeated identical loads) and linearity (deviation from a straight line across the range) taken together.

    How fast does the PM-200 ship, and can it be repaired?

    In-stock units ship the same or next business day from Cypress, Texas, with free ground delivery to the mainland USA in 1–5 business days. Out of warranty, the common failure points on this class of instrument — load cell or force-restoration coil, keypad membrane, display ribbon and power supply — are all serviceable, and we perform the repair plus a full recalibration in-house rather than returning the unit to the manufacturer.

    PM-200 manual summary and troubleshooting

    Unpacking and installation

    Site away from draughts and above a heat-tolerant surface. Fit the pan handler and a fresh aluminium sample pan, and confirm the exhaust above the unit is unobstructed.

    Calibration and verification

    Verify the weighing side with a certified weight, then verify the temperature side with a temperature calibration kit. Fix the sample mass, drying temperature and switch-off criterion in a written method; results are only comparable within the same method.

    Error codes and symptoms

    SymptomCause and remedy
    Result varies between operatorsSample mass or spreading differs. Standardise mass and spread the sample evenly and thinly.
    Drying never endsSwitch-off criterion too tight for the sample. Loosen the criterion or use a timed method.
    Unstable / flashing readingAir movement, vibration or an unlevel surface. Move away from HVAC vents and doors, level the bubble, and let the load settle before reading.
    Over-capacity indication (E, EEEE, or ---- )Applied load exceeds capacity, or the platter is fouled. Remove the load, check nothing is trapped under the platter, then re-zero.

    Care and re-verification

    Replace the halogen lamp per the manual's hour rating, keep the heating chamber clean, and never dry solvents or flammables.

    Need the printed PDF? The official RADWAG Balances and Scales manual library is at RADWAG Balances and Scales documentation, or call 832-290-3120 and we will email the exact revision for your serial number. See also the manual & documentation library.

    Customer reviews for PM-200

    No reviews yet. Be the first to review this product.